Receiving a Radioactive Package (10 CFR 20.1906)

Receiving a radioactive package: the monitoring deadline and the limits.

Open the tool → Runs in your browser. Data leaves only if you deliberately send a problem report from the interactive tool.

Example

Emitter type
Beta or gamma emitter, or a low toxicity alpha emitter (240 dpm/cm²)
Wipe: net counts per minute
2000 cpm
Wipe: counter efficiency
10 %
Wipe: area wiped
300 cm²

Result: Removable contamination 667 dpm/cm², above the 240 dpm/cm² limit.

The tool opens with these values already filled in. Replace them with your own.

What you enter

  • Date and time received (YYYY-MM-DDTHH:MM)
  • Received during normal working hours
  • Start of the next working day, if received after hours (YYYY-MM-DDTHH:MM)
  • Bears a White I, Yellow II or Yellow III label
  • Evidence of degradation (crushed, wet or damaged)
  • Holds only a gas or special form material
  • Contents exceed a Type A quantity
  • Emitter type (for the wipe limit)
The other 6 fields
  • Wipe: net counts per minute
  • Wipe: counter efficiency
  • Wipe: area wiped
  • Wipe efficiency, if measured (the rule allows 10% to be assumed)
  • Dose rate at the package surface
  • Dose rate at 1 meter
How this is calculated

Citation

10 CFR 20.1906(b), (c) and (d): which packages are monitored, within 3 hours, and when to notify the carrier and the NRC. 49 CFR 173.443 Table 9 (through 10 CFR 71.87(i)): 240 and 24 dpm/cm2. 10 CFR 71.47(a): 200 mrem/h and a transport index of 10. Read the sources: 10 CFR 20.1906 ↗, 49 CFR 173.443 ↗, 10 CFR 71.47 ↗

A reference and educational tool. Not medical, legal, or financial advice, and not a substitute for clinician judgment.