Sealed Source Leak Test Result and Next Due Date (10 CFR 35.67)

Sealed source leak test: the wipe activity and whether the source is leaking.

Open the tool → Runs in your browser. Data leaves only if you deliberately send a problem report from the interactive tool.

Example

Exemption under 35.67
None of these: a leak test is required
Net counts per minute on the wipe
500 cpm
Counter efficiency
10 %
Date of the leak test
2026-10-01

Result: 0.00225 µCi (83.3 Bq) of removable contamination: below the 0.005 µCi (185 Bq) level. Next test due by April 1, 2027.

The tool opens with these values already filled in. Replace them with your own.

What you enter

  • Exemption under 35.67(f)
  • Net counts per minute on the wipe
  • Counter efficiency
  • Date of the leak test (YYYY-MM-DD, optional)
How this is calculated

Citation

10 CFR 35.67(b), (c), (e) and (f): leak test at intervals not to exceed 6 months; 185 Bq (0.005 microcurie) or more of removable contamination is a leaking source; exemptions. 10 CFR 35.3067: report within 5 days. Read the sources: 10 CFR 35.67 ↗, 10 CFR 35.3067 ↗