Drug Clearance by CRRT (Flow and Sieving Coefficient)

Drug clearance by continuous renal replacement therapy from the effluent flows and a sieving or saturation coefficient.

Open the calculator → Runs in your browser. Data leaves only if you deliberately send a problem report from the interactive tool.

Example

Mode: hemofiltration, hemodialysis or hemodiafiltration
Hemofiltration (CVVH)
Flow unit
mL/hour
Ultrafiltrate flow
2000
Sieving coefficient
0.8
Patient's other clearance
10 mL/min

Result: CRRT clears 26.67 mL/min (1.6 L/hour), 72.73% of a total of 36.67 mL/min.

The tool opens with these values already filled in. Replace them with your own.

What you enter

  • Mode: hemofiltration, hemodialysis or hemodiafiltration
  • Flow unit: mL/hour (mL-h), mL/min (mL-min) or L/hour (L-h)
  • Ultrafiltrate flow (filtration modes)
  • Sieving coefficient (filtration modes)
  • Dialysate flow (dialysis modes)
  • Saturation coefficient (dialysis modes)
  • Plasma flow (with fluid before the filter)
  • Pre-filter flow
  • Patient's other clearance (optional)
How this is calculated

Citation

Li L, et al. Front Pharmacol. 2020;11:786 (convective clearance = sieving coefficient × ultrafiltrate flow; diffusive = saturation coefficient × dialysate flow); Corona A, et al. Antibiotics. 2022;11:1811 (hemofiltration with fluid given before the filter). Read the sources: Li 2020 ↗, Corona 2022 ↗

A reference and educational tool. Not medical, legal, or financial advice, and not a substitute for clinician judgment.