Pulse Oximeter Accuracy (Arms) From Paired Readings

Pulse oximeter accuracy from paired SpO2 and co-oximeter readings: Arms, bias and Bland-Altman limits for repeated measures.

Open the calculator → Runs in your browser. Data leaves only if you deliberately send a problem report from the interactive tool.

Example

Sensor type
Transmittance (wrap or clip)
Pairs, one per line
a, 91, 90 a, 93, 90 b, 92, 90 b, 94, 90 c, 90, 90 c, 92, 90

Result: Arms 2.38% over all pairs (within the 2013 guidance's typical 3.0% for this sensor)

The tool opens with these values already filled in. Replace them with your own.

What you enter

  • Sensor type
  • Pairs, one per line: subject, SpO2, SaO2
  • Subjects: subject, dark yes/no, MST 1-10, ITA, sex
How this is calculated

Citation

FDA, Pulse Oximeters - Premarket Notification Submissions [510(k)s] (2013, Table 3) and the January 2025 draft guidance (Arms definition, enrollment and bias criteria); Bland JM, Altman DG. J Biopharm Stat 2007;17:571-582. Read the sources: FDA 2013 guidance ↗, FDA 2025 draft ↗, Bland and Altman 2007 ↗