CT Effective Dose (DLP × k)

CT effective dose (AAPM Report 96): effective dose (mSv) = DLP × region-specific k conversion coefficient.

Open the calculator → Runs in your browser. Data leaves only if you deliberately send a problem report from the interactive tool.

Example

Dose-length product
400
Body region
Chest (k 0.014)

Result: Effective dose 5.6 mSv (chest, k 0.014).

The tool opens with these values already filled in. Replace them with your own.

What the result means

Caveat
A population estimate (ICRP-60 weighting), not patient-specific organ dosimetry; pediatric factors differ and are not used here.

What you enter

  • Dose-length product (mGy·cm)
  • Body region
How this is calculated
Formula
Effective dose (mSv) = DLP (mGy·cm) × region k. Adult k: head 0.0021, head-neck 0.0031, neck 0.0059, chest 0.014, abdomen/pelvis 0.015.

AAPM Report 96. The Measurement, Reporting, and Management of Radiation Dose in CT. American Association of Physicists in Medicine, 2008 (DLP × region conversion coefficient, EUR 16262 / ICRP-60 weighting). Read the source ↗

A reference and educational tool. Not medical, legal, or financial advice, and not a substitute for clinician judgment.

Browse all tools

Built by Clay Good. Source on GitHub.