Calibration Verification: Recovery and Reportable Range (CLIA)

Calibration verification recovery at each level against the laboratory's limit, with reportable range coverage.

Open the calculator → Runs in your browser. Data leaves only if you deliberately send a problem report from the interactive tool.

Example

One level per line
10, 10.2 100, 98 250, 254 500, 512
Laboratory's acceptable recovery limit
3
Reportable range, low end
10
Reportable range, high end
500

Result: All 4 levels within ±3% recovery.

The tool opens with these values already filled in. Replace them with your own.

What you enter

  • One level per line: assigned value, measured value
  • Laboratory's acceptable recovery limit (± %)
  • Reportable range, low end (optional)
  • Reportable range, high end (optional)
How this is calculated

Citation

42 CFR 493.1255(b): calibration verification with the laboratory's own limits, at a minimal (or zero), a mid-point and a maximum value, at least every 6 months. Recovery and the least-squares line are stated arithmetic. Read the source ↗

A reference and educational tool. Not medical, legal, or financial advice, and not a substitute for clinician judgment.

Browse all tools

Built by Clay Good. Source on GitHub.